The following is based on a poster presented at the 1999 ASLO Meeting in Santa Fe, NM:

AN IN SITU FLOW CYTOMETER FOR THE OPTICAL ANALYSIS OF INDIVIDUAL PARTICLES IN SEAWATER

Olson, R.J., Woods Hole Oceanographic Institution, Woods Hole, MA 02543, USA
Sosik, H.M., Woods Hole Oceanographic Institution, Woods Hole, MA 02543, USA

Abstract

Flow cytometry has proved a valuable tool for the analysis of phytoplankton and other suspended particles because of its speed and quantitative measurements, but the method’s oceanographic application has been limited by the need to take discrete water samples for analysis on board ship or in the laboratory.  For this reason, we have developed an in situ flow cytometer, which can operate unattended.  This instrument differs from conventional flow cytometers in that it uses a simple ducted flow of seawater through the flow cell, rather than hydrodynamically focussing a stream of sample seawater by injecting it into a sheath of particle-free fluid.  The new instrument defines a sensing region in the center of a diode-pumped 532 nm laser beam, based on the intersection of 2 diode lasers; only signals from those particles which pass through all 3 beams are accepted.  This approach should be less susceptible than conventional flow cytometry to problems stemming from flow disturbances, and hence more amenable to unattended long-term operation. Tests in the laboratory and off the WHOI dock indicate that sensitivity is comparable to that of conventional flow cytometers.  Design details and initial results will be presented.

Approach


(click on image for better resolution)
Fig. 1.  Schema of the optical and detection systems.  For simplicity, not all lenses, dichroic mirrors and detectors are shown.  In the actual instrument photomultipliers measuring scattering from the two IR lasers, forward and side scattering from the green laser, and green-excited red and orange fluorescence.


(click on image for better resolution)
Fig. 2.   Cross-sectional view of the flow cell.  The intersection of two IR lasers defines a sensing region in the center of a third, 532 nm, laser beam.  All three beams are in the same plane, with particles flowing upward through the flow cell channel.  Data will be acquired only from particles passing through all three beams simultaneously, ensuring that only signals from the central, uniform part of the green beam will be collected.
 
 

(click on image for better resolution)
Fig. 3.  Snapshot of signals from 3 beads passing through through the flow cell.  Only 1 of these passed through both IR beams as well as the green beam and was acquired.
 
 


(click on image for better resolution)
Fig. 4.  Signals from particles in a seawater sample recorded over 1 ms.  Of several particles observed during this time, only 1 passed through both IR beams as well as the green beam and was acquired (arrow).
 


(click on image for better resolution)
Fig. 5.  Block diagram of the flow cytometer electronics.
 
 


(click on image for better resolution)
Fig. 6.   Integration of photomultiplier signals.  Each amplified signal is independently controlled, with its own threshold to start and stop integration.
 

Results


(click on image for better resolution)
Fig. 7.  In situ operation off the WHOI dock.  The first version of the instrument was pressure tested to 130 m and completed the initial phase of field testing off the WHOI dock in January 1998.  After aligning the laser beams in the laboratory using uniform plastic microbeads suspended in a bucket of seawater that was circulating through the instrument (left panel), the flow cytometer was transported to the dock and lowered into the water on a cable.  After operating for a few minutes, a suspension of 5.2 mm beads was injected into the water near the sample intake.  The “in situ” beads had scattering and fluorescence signals identical to those measured in the laboratory, indicating that alignment was stable (right panel).  We are especially encouraged by this result since it occurred despite an abrupt temperature shift of 25 oC.
 


(click on image for better resolution)
Fig. 8.  Flow cytometric signature of particles from a sample from the WHOI dock.   Dynamic range was increased compared to the in situ dock test, by splitting each signal for simultaneous amplification at 2 different gain settings.  In this test, the gains are offset by 25-fold.  When we replace the current PMT’s by new ones with improved signal:noise characteristics, this factor can be increased.  Dynamic range of more than 4 decades should be achievable with this approach (for comparison, the modified EPICS, using 2 detectors and log amps for each parameter measured, has a range of about 5 decades).
 

 Interpretation
 
Calibration of particle concentration is necessary because the number of particles detected depends on the volume flow rate through the sensing region defined by the intersection of the two IR laser beams.  This volume flow rate will depend on both the rate of pumping of sample through the instrument (by the adjustable MicroPump), and on the size of the  laser spots, which can be adjusted to vary the sensitivity.

As a consequence of the Gaussian intensity profiles of the IR beams, larger particles are more likely to be detected than small particles: A small particle must pass through the central, most intense part of the IR beam to produce a signal above threshold, while a larger particle may do so even if it passes through the outer edge of the beam.  This means that a size-dependent correction is necessary to obtain particle concentrations.  We can empirically derive a correction algorithm by analyzing known mixtures of particles, as illustrated here for beads from 0.5 to 6 mm in diameter.

This size dependence of sampling efficiency can be viewed as an advantage (when calibrated) because in most cases flow cytometric analyses of phytoplankton are limited by the relative rarity of large phytoplankton, which are usually far less abundant than small cells.  With this approach, the data collected will no longer be determined by the numbers of the smallest cells, but will be more closely related to biomass.
 
 


(click on image for better resolution)
Fig. 9.  Detection efficiency as a function of particle size (as indicated by green beam side scattering).  A mixture of 5 kinds of beads was analyzed by “conventional” flow cytometry (Coulter EPICS modified for large dynamic range, with a 488 nm argon ion laser) and by the submersible flow cytometer in the laboratory.
 
 


(click on image for better resolution)
Fig. 10.  Side scattering signals from the bead mixture.  The size-dependent correction was applied to the raw data from the in situ flow cytometer.
 
 

 
(click on image for better resolution)
Fig. 11.  Analysis of phytoplankton cultures.  Cultures of Dunaliella tertiolecta (8 mm diameter) and Nannochloris sp. (2-3 mm diameter) were analyzed with the conventional (EPICS) and submersible flow cytometers.  The submersible’s raw data overestimates the concentration of the larger cells relative to the smaller ones.  After application of the correction derived from the bead mixture described earlier, the data is in good agreement with that of the conventional flow cytometer (see Fig. 12).
 

 
(click on image for better resolution)
Fig. 12.  Distribution of side scattering signals from phytoplankton culture samples.  The size-dependent correction applied to the raw data from the in situ flow cytometer (left panel) brings the results close to those from the conventional flow cytometer (right panel).
 

(click on image for better resolution)
Fig. 13.  Submersible flow cytometer in its frame, with battery case below).


(click on image for better resolution)
Fig. 14.  Housing removed to show optical bench and electronics.


(click on image for better resolution)
Fig. 15.  Top view of optical bench.
 
 

Future Directions

 Over the next few months, we will continue testing the instrument off the WHOI dock to evaluate laser beam and flow stability.  We also need to implement mechanisms for injecting bead mixtures for calibration and cleaning solutions to prevent fouling.  We then plan to deploy the in situ flow cytometer at the LEO-15 mooring (which can supply power and transmit data to shore) off New Jersey, to examine time series data of phytoplankton and other particles in the coastal environment.  We also envision a ship-based version of the flow cytometer to continously analyze phytoplankton in samples from the ship’s intake or from sample water pumped from a SeaSoar vehicle.
 
 

Go back to Olson Lab Homepage