General Introduction
Two new Finnigan MAT253 mass spectrometer systems for the analysis of
calcium carbonate samples were purchased by the Paleoceanography
research group at WHOI in 2003 and 2004. They were installed by
Finnigan in March 2004 and have been in operation since July 2004.
The purpose of this page is to document the procedures and capabilities
of these instruments. During the interval between March and July
2004, we analyzed more than 1200 weighed carbonate standards (NBS18,
NBS19, Carrara, B1 and AtlantisII) and our facility has analyzed over
145,000 samples for 95 researchers from 43 different Institutions
and Universities worldwide.
The precision of 600 NBS19 analyses (all sizes
from 15 to 300µg) was ±0.08 for
δ18O and ±0.03 for
δ13O.
Small sample precision (<20µg and under all laboratory
conditions) is ±0.10 for
δ18O
and ±0.05 for
δ13C.
Based on the standard analyses, we have developed an empirically-based
calibration of our measurements to PDB which includes corrections
for the following analytical bias:
- Differences in the mean value of our standards through time
- These differences are caused by variations in our analytical
procedures (length of reaction, trap temperatures during reactions
and transfers, duration of transfers) and caused by failures
of some important components during sample analyses (traps that
did not achieve cold enough temperatures for compete gas transfers).
The magnitude of these variations is 0.15
o/oo
for δ18O
and 0.04 for δ13C.
After correcting for these biases, the precision for all analyses
of NBS19 is ±0.07 for
δ18O
and ±0.03 for
δ13C.
For samples between 10 and 20 µg, the precision is poorer, but
still an acceptable ±0.117 for
δ18O
and ±0.05 for
δ13C.
|
Download plot of daily samples

last updated May 2008
|
|